• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPORT ISP IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, 8F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPORT ISP IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, 8F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Catalog Number 4708060
Medical Device Problem Codes Device Appears to Trigger Rejection (1524); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Bacterial Infection (1735); Unspecified Infection (1930); Thrombosis/Thrombus (4440)
Date of Event 10/26/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: the physical device was not returned for evaluation.No medical records were provided for review.Therefore, the investigation is inconclusive for the reported infection and thrombosis as no objective evidence was provided for review.Furthermore, the clinical conditions alleged in the complaint cannot be confirmed.Based upon the available information, the definitive root cause is unknown.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported through litigation process that sometime later port placement procedure, the patient developed with unspecified infection and thrombosis.The current status of the patient was unknown.
 
Additional Manufacturer Narrative
H11: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: although the physical device was not returned for evaluation, a medical record was provided for review.Medical record alleges that, the patient underwent an implantable port placement procedure via right subclavian vein using a powerport for iv fluid delivery as part of postural orthostatic tachycardia syndrome treatment.Approximately, five months and fourteen days later the patient was reported with staphylococcus aureus infection.Subsequently, the patient was planned for port removal.A previous incision was performed, and the pocket was entered.There was no purulent or turbid fluid noted.The pocket reservoir itself was cultured and submitted for gram stain and culture.The port catheter was dissected free from the fibrin sheath, and the fibrous sheath around the catheter was incised, and the catheter was removed.The fibrin sheath was oversewn with suture.The catheter tip was submitted for culture, and the reservoir was sent for a definitive culture.Therefore, the investigation is confirmed for the reported device appeared to trigger rejection issue.Furthermore, the clinical conditions alleged in the complaint can be confirmed.Based upon the available information, the definitive root cause is unknown.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.B3, b5, g3, h6 (device, component, method, result) section a through f: the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported through the litigation process that, on (b)(6)2017, a patient underwent port placement via the right subclavian vein for intravenous fluid delivery related to postural orthostatic tachycardia syndrome.Approximately five months and fourteen days later, on (b)(6) 2017, the patient was diagnosed with bacterial infection.Additionally, it was reported that the patient was diagnosed with thrombosis.Subsequently, the port was removed.However, the current status of the patient was unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPORT ISP IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, 8F
Common Device Name
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key22138821
Report Number3006260740-2025-03614
Device Sequence Number12820176
Product Code LJT
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K072215
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2018
Device Catalogue Number4708060
Device Lot NumberREBP1217
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/22/2025
Supplement Date Received by Manufacturer06/09/2026
Initial Report FDA Received Date06/04/2025
Supplement Report FDA Received Date06/22/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age30 YR
Patient SexMale
-
-