A physician reported that the actuation, cutting, and aspiration failure of a probe occurred during vitrectomy surgery.The surgery was completed after replacing the product with another one.The involved eye and the patient identifier are not available.There was no patient impact.
|
|
Additional information provided in d.9., h.3., h.6.And h.11.One opened probe, with tip protector, in a tray with other items was received for the report.The sample was visually inspected and found to be nonconforming with orange/brown foreign material on the welded cap and white/clear foreign material was observed on the port face, inside of port and on body needle.The sample was then functionally tested for actuation, aspiration and cut.The sample was found to be conforming for actuation, nonconforming for aspiration and nonconforming for cut.The probe was disassembled and the components inspected.Excessive wear observed on inner cutter when compared to the degree of wear based on continuous actuation of the probe visual standard photos.Gouge marks at cutting edge, bend area, and several other areas on the inner cutter.The sample was tested with a syringe and resistance was felt.A wire was inserted through the aspiration path.The sample was retested with a syringe, and no resistance was felt.The sample was retested for aspiration with the probe driver and was able to aspirate.The initial aspiration test failed due to an interference in the aspiration path and once the interference was removed the probe was able to aspirate.A review of the device history record traceable to the lot number obtained from the device¿s radio frequency identification (rfid) tag, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The investigation conducted a non-conformance review of the reported lot number.No deviations were identified, and all corresponding production release specifications defined in the device master record were met.Based on the evaluation of the information and materials received, the investigation was unable to identify the root cause or origin of the reported event.Additionally, no manufacturing-related deficiencies were found that potentially could have contributed to the complaint.The complaint evaluation confirms the probe had an aspiration and cut failure.The root cause of the aspiration failure is an interference in the aspiration path once the interference was removed the probe was able to aspirate.The complaint evaluation also confirms that probe had a cut failure.The root cause for the poor cutting is the observed gouge marks.The probe was disassembled, and the components inspected.The wear observed on inner cutter when compared to the degree of wear based on continuous actuation of the probe visual standard photos is consistent with excessive use classification.Excess usage of the probe will wear and damage the inner cutter such that the cutter function becomes poor, bent, or does not function at all.Per the direction for use (dfu), the vitrectomy probe is verified for 20 minutes of use at maximum cut speed.The observed cut and aspiration failures were due to excessive use of the probe by the user.Per the dfu, the vitrectomy probe is verified for 20 minutes of use at maximum cut speed.A nonconformance investigation is currently in progress to investigate the trend identified for probes with would not cut or actuate issues.All probes are 100% visually inspected and tested for actuation, aspiration, and cut during manufacturing.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.Complaint data for all company products is reviewed monthly to monitor for adverse trends.During the last review, adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
|