| Catalog Number |
51-104150 |
| Medical Device Problem Code |
Manufacturing, Packaging or Shipping Problem (2975)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/09/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that while inspecting the packaging, the sterile packaging was found damaged.There was no patient involvement.There is no further information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).D10: 51-104120 item name tprlc 133 t1 pps ho 12x144mm lot# 7004917; d10: 51-109050 item name tloc 133 mp sp t1 pps ho 5x95 lot # 7224177; d10: 51-145170 item name tprlc xr mp t1 pps 17x119mm lot # 7338366; d10: 51-104140 item name tprlc 133 t1 pps ho 14x148mm lot # 6289615; d10: 51-107160 item name tprlc 133 mp type1 pps ho 16.0 lot # 7124269.G2: foreign: japan.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
No further information is available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated: d9; g3; h2; h6; h10.Visual evaluation of the returned product/provided photos identified damage to the sterile packaging (blister).Sterility has not been compromised.The reported event did not occur in an operating room or as part of a medical procedure; medical records are not available for review.The root cause of the reported event can be attributed to transit damage and a packaging design issue.The reported event has been confirmed by evaluation of the provided photos/returned product.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Search Alerts/Recalls
|