• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT THEON PERICARDIAL BIOPROSTHESIS; REPLACEMENT HEART VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT THEON PERICARDIAL BIOPROSTHESIS; REPLACEMENT HEART VALVE Back to Search Results
Model Number 2700TFX
Medical Device Problem Code Intravalvular regurgitation (4068)
Health Effect - Clinical Code Dyspnea (1816)
Date of Event 05/12/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H11: additional manufacturer narrative: the investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event or Problem Description
It was reported that a patient with a 25mm 2700tfx aortic valve underwent a valve-in-valve procedure after an implant duration of 7 years, 11 months due to regurgitation.The patient presented with shortness of breath.The tavr was successfully performed with a 23mm 9755rsl valve.The patient was in stable condition post procedure and discharged on pod #1.
 
Additional Manufacturer Narrative
H11: additional manufacturer narrative: updated: g3, g6, h2, h6: (type of investigation, investigation findings, investigation conclusions).Regurgitation is a common manifestation of bioprosthetic valve and valved conduit dysfunction.Common intraoperative factors that may cause or contribute to acute post-procedural regurgitation include device distortion; device interaction with patient anatomy or other devices; leaflet damage, and incorrectly sized valve seated.Non-structural valve dysfunction (nsvd), endocarditis and/or thrombosis may also cause or contribute to acute post-procedural regurgitation.Complaints reported with regurgitation as the primary complaint code are not typically attributed to manufacturing-related nonconformances.A device history record (dhr) review was performed, and no relevant non-conformances were identified.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.Based on the information available, the most likely root cause is patient, including chronic kidney disease, hyperlipidemia, diabetes mellitus.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT THEON PERICARDIAL BIOPROSTHESIS
Common Device Name
REPLACEMENT HEART VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
mle fl2 - office m2013
irvine, CA 92614
9492506615
MDR Report Key22142750
Report Number2015691-2025-04407
Device Sequence Number15234546
Product Code DYE
UDI-Device Identifier00690103194043
UDI-Public(01)00690103194043(17)200502
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P860057
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/01/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/02/2020
Device Model Number2700TFX
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/12/2025
Supplement Date Received by Manufacturer07/01/2025
Initial Report FDA Received Date06/04/2025
Supplement Report FDA Received Date07/01/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/03/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention; Life Threatening;
Patient Age72 YR
Patient SexFemale
-
-