H11: additional manufacturer narrative: the investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
|
H11: additional manufacturer narrative: updated: g3, g6, h2, h6: (type of investigation, investigation findings, investigation conclusions).Regurgitation is a common manifestation of bioprosthetic valve and valved conduit dysfunction.Common intraoperative factors that may cause or contribute to acute post-procedural regurgitation include device distortion; device interaction with patient anatomy or other devices; leaflet damage, and incorrectly sized valve seated.Non-structural valve dysfunction (nsvd), endocarditis and/or thrombosis may also cause or contribute to acute post-procedural regurgitation.Complaints reported with regurgitation as the primary complaint code are not typically attributed to manufacturing-related nonconformances.A device history record (dhr) review was performed, and no relevant non-conformances were identified.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.Based on the information available, the most likely root cause is patient, including chronic kidney disease, hyperlipidemia, diabetes mellitus.
|