| Catalog Number |
MZ9226 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/09/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that bd maxzero microbore extension set was damaged the following information was received by the initial reporter with the following verbatim, it was reported by customer that iv syringe tubing with filter cracked and leaking.
|
| |
|
Additional Manufacturer Narrative
|
|
One sample model mz9226 was returned for investigation.The set was examined for defects and abnormalities.No defects or abnormalities were observed.The set was attached to a 10 ml bd syringe and flushed with water.A leakage was observed come from the vent of the filter.The customer complaint was verified.Additional air under water test was performed.The set was pressurized with about 10psi, and the vent was submerged under water.There was a steady stream of air bubbles coming from the filter vent.This would indicate that there is no issue with damage/ holes in the membrane.The reason for the membrane leaking is that the hydrophobicity of the membrane must have been compromised during use and not during the manufacturing process.A device history record review could not be performed because a lot number was not provided by the customer.
|
| |
|