• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAUSCH + LOMB EASY-LOAD LENS DELIVERY SYSTEM; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAUSCH + LOMB EASY-LOAD LENS DELIVERY SYSTEM; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL) Back to Search Results
Model Number EZ-24
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Capsular Bag Tear (2639)
Date of Event 05/05/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Although requested the device was not returned for evaluation.A review of the device history record (dhr) did not identify any anomalies or nonconformities that could be related to this event.The lot history, risk analysis and directions for use review were considered acceptable, with the product performing within anticipated rates.Based on the available information, user related factors such as loading or handling techniques and/or procedural factors such as lens and inserter interaction may have contributed to the event.
 
Event or Problem Description
A surgical center reported that during the implantation of an intraocular lens (iol) into the patients left eye a capsular tear occurred upon ejection of the lens.When the iol was inserted into the eye, the doctor noticed the leading haptic was over the optic.As he tried to fix the lens, a tear was created in the capsular bag.The lens was explanted, an anterior vitrectomy was performed, and a backup lens was successfully placed.Per the report the patient is expected to make a full visual recovery.
 
Additional Manufacturer Narrative
The reporter returned 4 open injectors with no indication of which injector was the used in the surgery.We evaluated all 4 injectors and results are as follows: injector #1.The injector was received loose in an opened pouch.Microscopic examination was performed and found no visible damage.The plunger is advanced into the cartridge.Injector #2.The injector was received loose in an opened pouch.Microscopic examination was performed and found no visible damage.The plunger is stuck due to dried solution.Injector #3.The injector was received loose in an opened pouch.Microscopic examination was performed and found no visible damage.The plunger is advanced into the loading deck.Injector #4.The injector was received loose in an opened pouch.Microscopic examination was performed and found no visible damage.The plunger is advanced into the loading deck.A review of the device history record (dhr) did not identify any anomalies or nonconformities that could be related to this event.The lot history, risk analysis and directions for use review were considered acceptable, with the product performing within anticipated rates.Based on the available information, user related factors such as loading or handling techniques and/or procedural factors such as lens and inserter interaction may have contributed to the event.Additional information: d9, g1, g3, h2, h3, h6, h11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASY-LOAD LENS DELIVERY SYSTEM
Common Device Name
FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)
Manufacturer (Section D)
BAUSCH + LOMB
1400 north goodman street
rochester NY 14609
Manufacturer (Section G)
BAUSCH + LOMB
21 north park place blvd.
clearwater FL 33759
Manufacturer Contact
shayan habibi
21 north park place blvd.
clearwater, FL 33759
7277246600
MDR Report Key22145321
Report Number0001313525-2025-00121
Device Sequence Number15585056
Product Code MSS
UDI-Device Identifier00757770529032
UDI-Public(01)00757770529032(17)260612(10)H2416301
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K132481
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 05/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/12/2026
Device Model NumberEZ-24
Device Lot NumberH2416301
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/06/2025
Supplement Date Received by Manufacturer06/13/2025
Initial Report FDA Received Date06/04/2025
Supplement Report FDA Received Date07/11/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/12/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SOFPORT AO INTRAOCULAR LENS.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age77 YR
Patient SexFemale
-
-