MAUDE Adverse Event Report: BIOTRONIK AG, BUELACH, SWITZERLAND PASSEO-18 2/60/150; PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK AG, BUELACH, SWITZERLAND PASSEO-18 2/60/150; PERIPHERAL DILATATION CATHETER Back to Search Results
Model Number 366119
Medical Device Problem Codes Material Puncture/Hole (1504); Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/16/2025
Type of Reportable Event Malfunction
Event or Problem Description
The passeo-18 peripheral balloon catheter was selected for dilatation of the mid peroneal artery with 80 percent stenosis degree.The first balloon ruptured at 6 atm (reported separately).Therefore, the second passeo-18 was used but ruptured as well at the first inflation.The balloon was removed from patients body.
 
Additional Manufacturer Narrative
The returned device was subjected to a technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The technical investigation revealed a fine jet of water emerging from the center of the balloon when attempting inflation.Microscopic analysis of the balloon surface showed a small pinhole with longitudinal orientation direct at the distal end of the distal x-ray marker.At the origin of the pinhole a deep scratch was observed which has likely been caused by a hard, sharp-edged object such as e.G., anatomical structure.Review of the production documentation confirmed that the device was manufactured according to specifications and passed all in-process and final inspections.In addition to visual inspections each device is tested for air tightness by means of a helium leak test.We can therefore confirm that the device was delivered in a leak-proof condition.Based on the conducted investigations, no manufacturing or material related root cause could be determined.The root cause is most likely related to the patient¿s anatomy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PASSEO-18 2/60/150
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key22146882
Report Number1028232-2025-02731
Device Sequence Number8469128
Product Code LIT
UDI-Device Identifier07640130414617
UDI-Public(01)07640130414617(17)2710
Combination Product (Y/N)N
Initial Reporter CountryMY
PMA/510(K) Number
K072765
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number366119
Device Catalogue Number366119
Device Lot Number09242301
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/22/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/03/2025
Supplement Date Received by Manufacturer06/19/2025
Initial Report FDA Received Date06/04/2025
Supplement Report FDA Received Date06/19/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/02/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-