| Model Number |
SONATA |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/04/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2025, the parent reported that the recipient had not been hearing sounds with the device for about two weeks.The recipient has been re-implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: based on the received information, a damage to the active electrode due to excessive mechanical stress appears very likely.However, to determine an exact root cause device investigation would be necessary.The concerned device was explanted but has not been received for investigation yet.This is a combined initial and final report.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2025, the parent reported that the recipient had not been hearing sounds with the device for about two weeks.The recipient has been re-implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: overload fractures in the active electrode which are consistent with an external mechanical impact were determined to be the root cause of device failure.The problems given in the recipient report appear to match the damage found.This is a final report.
|
| |
|