Model Number 43103
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/02/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Medtronic received information regarding a catheter.It was reported that the catheter was leaking.The procedure was completed with back up products.The patient's status at the time of the report was alive-no injury.
 
Additional Manufacturer Narrative
H3.Product analysis determined that the visual inspection confirmed the catheter has been damaged/cut.The returned product did not meet the requirements for patency and leak testing.All catheters are 100% inspected at the time of manufacture.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIAC/PERITONEAL CATHETER
Common Device Name
SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
Manufacturer (Section D)
MEDTRONIC DOMINICANA
zona franca de san isidro
carretera san isidro, km 17
santo domingo 11500
DR  11500
Manufacturer (Section G)
MEDTRONIC DOMINICANA
zona franca de san isidro
carretera san isidro, km 17
santo domingo 11500
DR   11500
Manufacturer Contact
justin ellis
4620 n. beach st.
fort worth, TX 76137
7635265677
MDR Report Key22149018
Report Number9612501-2025-01444
Device Sequence Number12878835
Product Code JXG
UDI-Device Identifier00763000028305
UDI-Public(01)00763000028305(17)300112(10)0230648713
Combination Product (Y/N)N
Initial Reporter CountryID
PMA/510(K) Number
K792005
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number43103
Device Catalogue Number43103
Device Lot Number0230648713
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/04/2025
Supplement Date Received by Manufacturer07/03/2025
Initial Report FDA Received Date06/05/2025
Supplement Report FDA Received Date07/07/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/07/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-