Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pulmonary Embolism (1498)
Date of Event 03/31/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D10: 42502006202, femur cemented cruciate retaining (cr) narrow right size 7, 66972253.42532006702, tibia cemented 5 degree stemmed right size d, 67064898.42522100410, articular surface medial congruent (mc) right 10 mm height use with tibia sizes c-d/cr femur sizes 6-7, 67017006.61911010, bone cement, rkf280.Customer has indicated that the product will not be returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the patient underwent a right total knee arthroplasty.Five days later, the patient was hospitalized due to a pulmonary embolism.Patient was discharged approximately 10 days later and was reported as resolved.All available information has been provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; d2; g1; g3; g6; h1; h2; h3; h6 the event cannot be confirmed.Product has not been evaluated as it has been determined the event is not related to device.No further actions will be initiated as a result of reported event.Root cause of the reported event is not device related.Procedural related complications are influenced by the type of surgery, patients pre-existing comorbidities, and perioperative management.Deep vein thrombosis (blood clot), or dvt, occurs when a blood clot forms in one of the deep veins of the body.This can happen if a vein becomes damaged or if the blood flow within a vein slows down or stops.Total joint patients are typically placed on medication post-operatively to prevent dvt formation.Despite prophylaxis, dvts can still develop which can then break free within the vessel and occlude or block the blood flow in the lungs, known as a pulmonary embolism (pe).As the complaint indicated, a post-operative complication occurred, and medical intervention was required for treatment.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONA ALL POLY PATELLA CEMENTED 29 MM DIAMETER
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer (Section G)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key22149133
Report Number0002648920-2025-00153
Device Sequence Number7638780
Product Code MBH
UDI-Device Identifier00889024643697
UDI-Public(01)00889024643697(17)290924(66927616)
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K172524
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/26/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42540000029
Device Lot Number66927616
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/13/2025
Supplement Date Received by Manufacturer08/29/2025
Initial Report FDA Received Date06/05/2025
Supplement Report FDA Received Date09/26/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/25/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Life Threatening; Hospitalization; Required Intervention;
Patient SexUnknown
-
-