|
Updated fields: h2, h10, h11.The sureform 60 reload associated with this report was not returned for failure analysis.Additional information: a concomitant product, the sureform 60 stapler used to fire the blue reloads, and another blue reload were received and failure analysis investigations were performed.These products were received in reference to mdr with mfr report number 2955842-2025-24052.The sureform 60 stapler instrument for failure analysis investigation.System logs showed firing failure and incomplete (hi torque) errors.The sureform 60 stapler instrument was placed and driven on an in-house system.The instrument was fully functional.No product issue was identified.The blue sureform 60 reload was returned in used condition.A visual inspection discovered that the stapler reload was partially fired.Further inspection revealed that, up to this firing interruption, all staples were fired, the shuttle was okay, and the shuttle knife had damage.The reload was found with a damaged blade inside the cartridge track.Further inspection revealed that the blade had mechanical indentations.A visual inspection did not show blade rotation.The reload was found to have a cartridge damage along the top surface.The surface showed bulges at the point where the knife stopped.Additionally, the stapler logs for this procedure were reviewed.The stapler instrument was installed on the system 5 times and fired 5 reloads (1 green, 4 blue) during the procedure.On installs 1, 3, and 4, the first clamp was successful, and the firings were each completed with no pauses for compression.On install 2, all clamps were successful, and the firing was completed with 1 pause for compression.A firing failure was observed on the fifth fire with a representative error.Another stapler instrument was installed to continue the procedure.
|