| Model Number |
KD-VC611Q-07301S |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/27/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the final investigation.The device was returned to olympus for inspection and the reported failure was not confirmed.However, the following reportable malfunction was identified during the device evaluation: the coated portion of cutting wire was torn.Based on the results of the investigation, and since the customer reported malfunction was not reproduced, a root cause could not be identified.Based on the results of the investigation, a root cause for the malfunction identified during the device evaluation could not be identified.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
No additional information was received from the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the cutting wire on the sphincterotome broke when it was tightened.The event occurred during an endoscopic retrograde cholangiopancreatography (ecrp).There were no reports of patient harm.
|
| |
|