It was reported that during a shoulder arthroscopy surgery, all 3 devices broke off during implantation.The tip of the device bended 90 degrees.Therefore, self-punching was not possible.The broken piece was retrieved from the patient.Per complaint reporter there was no harm for patient, operator or third party.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
|
|
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to improper bone prep; misaligned insertion and/or prying/leveraging the device during insertion.The dfu for this device, dfu-0087-13, gives the following precautions: pushlock and swivelock suture anchors only: insert the driver into the bone socket until the anchor body makes contact with the bone.Preview and adjust suture tension, if necessary.Tension will not increase during final advancement of the anchor body.Pushlock and swivelock suture anchors only: ensure that the anchor body is in full contact with the bone before advancing the anchor body into the prepared bone socket.Self-punching suture anchors only: insertion in very hard bone may require pre-punching a bone socket to avoid damage to the implant.Self-punching pushlock and swivelock suture anchors only: ensure that the angle of anchor insertion is perpendicular to the bone.
|