| Model Number |
72075-01 |
| Medical Device Problem Codes |
Device Alarm System (1012); Application Program Problem (2880)
|
| Health Effect - Clinical Codes |
Hypoglycemia (1912); Shaking/Tremors (2515); Convulsion/Seizure (4406)
|
| Date of Event |
06/02/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
An alarm issue was reported with the abbott diabetes care (adc) device in use with huawei p20 pro with android operating system version 10.The low glucose alarm did not sound and the customer was not alerted of changes in glucose level.As a result, the customer experienced symptoms described as shaking, freezing, a seizure, and unable to speak.The customer was unable to self-treat, requiring third-party administration of jubin sugar solution for treatment.There was no report of death or permanent impairment associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The customer experienced missing low glucose alarms with the reported application.Per the compatibility guide, use of the huawei p20 pro operating system 10 is incompatible with the reported application software version 3.6.4.This guide is available to the user on the websites where the product is launched.The most probable root causes associated with this failure mode are software/data corruption or misuse.However, mitigations are in place to reduce and prevent such issues.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.The device mfg date is not applicable.The date entered in section h4 is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation was performed for the reported complaint.The user reported missing low alarms.The reported configuration was not compatible with the customer's reported application.The latest revision of the compatibility guide was available to the customer on the abbott diabetes care website.As the compatibility guide is provided to the customer and the incompatible configurations were used, this complaint is not confirmed to use.This serves as a correction report.Section h11 (additional mfg narrative) was incorrectly updated in the previous report.Correction has been made.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
An alarm issue was reported with the abbott diabetes care (adc) device in use with huawei p20 pro with android operating system version 10 and unknown app version.The low glucose alarm did not sound and the customer was not alerted of changes in glucose level.As a result, the customer experienced symptoms described as shaking, freezing, a seizure, and unable to speak.The customer was unable to self-treat, requiring third-party administration of jubin sugar solution for treatment.There was no report of death or permanent impairment associated with this event.
|
| |
|
Search Alerts/Recalls
|