|
The investigation has determined that a non-reproducible, lower than expected glucose (glu) result was obtained from a single patient sample processed using vitros chemistry products glu slides lot 0036-3248-5857 on a vitros 5600 integrated system.A definitive assignable cause of the event could not be determined with the limited information provided.No historic quality control results obtained using vitros glu reagent lot 0036-3248-5857 were provided when requested, therefore, it is not possible to assess the performance of vitros glu reagent lot 0036-3248-5857 and a reagent related performance issue cannot be ruled out as contributing to the event.However, continual tracking and trending of complaints has not identified any signals that would point to a potential systemic issue with vitros glu reagent lot 0036-3248- 5857.An instrument cannot be completely ruled out as a contributing factor as precision testing to assess instrument performance was not performed by the customer when requested.Therefore, it cannot be confirmed that the instrument was operating as intended and unexpected instrument performance cannot be completely ruled out as contributing to the event.Pre-analytical sample processing could not be ruled out as a contributing factor as it was unknown if the customer was following the sample collection device manufacturer recommended centrifugation protocol.Improper pre-analytical sample handling could have contributed to the event.It is possible that cellular debris, due to poor sample preparation, was present in the affected sample, although this could not be confirmed.Since the vitros 5600 integrated system is not interfaced to e-connectivity, and the customer did not collect data log files, it was not possible to review sample metering profiles for the initial and repeat testing events to investigate the possibility of pre-analytical sample mix-up.Therefore, pre-analytical sample mix-up cannot be ruled out or confirmed as contributing to the event.
|