| Catalog Number |
ASK-04120-HF3 |
| Medical Device Problem Codes |
Dull, Blunt (2407); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/15/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported "the catheter is buckling on insertion.The tip of the 04120 is blunt and doesn't thread through the skin like the ra-04020." another catheter was used.There was no patient harm or injury.No medical intervention required.The patient's current condition is reported as "fine".Associated mdr numbers include: 9680794-2025-00467 and 9680794-2025-00478.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
|
| |
|
Event or Problem Description
|
|
It was reported "the catheter is buckling on insertion.The tip of the 04120 is blunt and doesn't thread through the skin like the ra-04020." another catheter was used.There was no patient harm or injury.No medical intervention required.The patient's current condition is reported as "fine".Associated mdr numbers include: 9680794-2025-00467 and 9680794-2025-00478.
|
| |
|