SYNTHES GMBH HOHMANN-STYLE ARM 183MM; FORCEPS Back to Search Results
Catalog Number 398.752
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that user had two items that broke postoperatively.The items won't click onto the handle anymore.This was noted in the sterile processing department.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: h4 h3, h4, h6: the product was returned to j&j medtech orthopaedics for evaluation.Visual inspection of the returned device found nothing indicative of a device nonconformance.Device has no traces of damage that could interfere in its connection to the handle.A dimensional inspection was not performed since it was not applicable to the complaint condition.A functional evaluation was partially performed using the internal mechanism of the hohmann-style arm 183mm and no defects were found that could interfere with the handle interaction.However, since mating device was not returned, we are not able to accurately replicate the reported condition.The overall complaint was not confirmed as the observed condition of the hohmann-style arm 183mm would have not contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review (dhr): part # 398.752 synthes lot # 2109116 supplier lot # 2109116 release to warehouse date: 01.Feb.2005 expiration date: na supplier : (b)(4).No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOHMANN-STYLE ARM 183MM
Common Device Name
FORCEPS
Manufacturer (Section D)
SYNTHES GMBH
luzernstrasse 19-21
zuchwil 4528
SZ  4528
Manufacturer (Section G)
BACHLER FEINTECH AG
postfach
hölstein 4434
SZ   4434
Manufacturer Contact
elba bello
1302 wright lane east
west chester, PA 19380
9083863534
MDR Report Key22153105
Report Number8030965-2025-05654
Device Sequence Number15798370
Product Code HTD
UDI-Device Identifier10886982201560
UDI-Public(01)10886982201560(10)2109116
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number398.752
Device Lot Number2109116
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/02/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/14/2025
Supplement Date Received by Manufacturer07/03/2025
Initial Report FDA Received Date06/05/2025
Supplement Report FDA Received Date07/04/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/01/2005
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-