| Catalog Number |
364391 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/13/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
B3.Date of event is unknown.The date received by manufacturer has been used for this field.E1: (b)(6).A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that before using one (1) bd preset¿ eclipse¿, foreign matter was observed on the plunger stopper.No adverse impact was reported regarding the patient or user.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: bd received three photos for investigation.Evaluation of the photos shows light brown colored foreign material on the top and side of the plunger stopper.A total of ten retained samples were visually inspected, and no foreign material was observed.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has been confirmed for the indicated failure mode: foreign matter - unknown.No definitive root cause could be established for the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported that before using one (1) bd preset¿ eclipse¿, foreign matter was observed on the plunger stopper.No adverse impact was reported regarding the patient or user.
|
| |
|
Additional Manufacturer Narrative
|
|
The following fields were updated due to additional information: d9: device available for evaluation: yes h.3 device eval by manufacturer? yes d9: returned to manufacturer on: 03-sep-2025 investigation summary bd received three photos and one sample for investigation.Evaluation of the photos and returned customer sample shows light brown colored foreign material on the top and side of the plunger stopper.Additionally, ten retained samples were visually inspected, and no foreign material was observed.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has been confirmed for the indicated failure mode: foreign matter - unknown.No definitive root cause could be established for the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported that before using one (1) bd preset¿ eclipse¿, foreign matter was observed on the plunger stopper.No adverse impact was reported regarding the patient or user.
|
| |
|
Search Alerts/Recalls
|