Event or Problem Description
|
|
It was reported that the subject device was broken.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the final investigation.Updated fields: d9, h2, h3, h6, h11.The device was returned to olympus for inspection, and the reported failure was confirmed.In addition, the following reportable malfunctions were identified during the device evaluation: fiber breakage, and material fragmentation under the lens.Based on the results of the investigation, a root cause could not be identified.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
No additional information received from the customer.
|
| |
|