|
MEDTRONIC SOFAMOR DANEK USA, INC CD HORIZON® SPINAL SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL
|
Back to Search Results |
|
| Model Number |
9010000849 |
| Medical Device Problem Code |
Mechanics Altered (2984)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/28/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3: product analysis: part # 9010000849; lot # rs17d004 visual and functional inspection revealed the inner shaft does not move when the locking lever is moved.The c clip that connects the adjusting mechanism to the inner shaft appears to have been overloaded and broken.It is possible to overload the clip when making tension adjustments while the rod is in place.It appears the c clip broke from excessive force.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Information was received from user facility via a manufacturer representative regarding a device used for spinal therapy.It was reported that the internal mechanism of the inserter, which tightens the instrument to hold the rod, was not functioning properly.There was no patient involved.
|
| |
|
Search Alerts/Recalls
|
|
|