• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER MANUFACTURING B.V. VERSYS HIP SYSTEM; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER MANUFACTURING B.V. VERSYS HIP SYSTEM; PROSTHESIS, HIP Back to Search Results
Catalog Number 00801803602
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Date of Event 05/31/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).G2: foreign ¿ australia.Suggested component code: mechanical (g04) ¿ head.The customer has indicated that the product will not be returned to zimmer biomet for investigation, as its location is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that approximately eight years post-implantation, the patient underwent a hip revision due to pain and difficulties with adls.All components were revised.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated: g3; h2; h3; h6; h10.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.Medical records/radiographs were provided.Review of the available records identified the following: right hip pain.Trouble with adl¿s due to hip pain.The lateral acetabular screw is extra-osseous.A definitive root cause cannot be determined.The complaint is confirmed based on the medical record findings.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSYS HIP SYSTEM
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key22157343
Report Number0002648920-2025-00154
Device Sequence Number7292513
Product Code LPH
UDI-Device Identifier00889024144729
UDI-Public(01)00889024144729(17)251031(10)63193583
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K953337
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2025
Device Catalogue Number00801803602
Device Lot Number63193583
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/13/2025
Supplement Date Received by Manufacturer08/27/2025
Initial Report FDA Received Date06/06/2025
Supplement Report FDA Received Date09/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/23/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexFemale
Patient Weight51 KG
-
-