Model Number CCW0T4
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Visual Impairment (2138); Fibrosis (3167); Posterior Capsule Opacification (4807)
Date of Event 05/15/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The product was not returned for analysis; the lens remains implanted.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
 
Event or Problem Description
A nurse reported that following an intraocular lens (iol) implant procedure, the patient had early posterior capsule opacification (pco).Patient was 3 weeks post-op with anterior and posterior capsule opaque with pin hole vision.Additional information has been requested, received and stated patient was 3 weeks post-op with anterior and posterior capsule contracted and opaque with pin hole vision.Patient issues were not resolved, scheduled for yag capsulotomy anterior and posterior.
 
Additional Manufacturer Narrative
Additional information provided in h.11.A global review was conducted to look at the reported incidence for each of the above outcomes over the past 3 years.All global complaints are evaluated monthly for adverse trending, and no adverse trends were identified for these events.File will be reopened if the sample is received.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON TORIC UV ABSORBING IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key22158794
Report Number1119421-2025-01618
Device Sequence Number15329891
Product Code MJP
UDI-Device Identifier00380652432917
UDI-Public(01)00380652432917(17)261016(11)221018(10)15511642
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 07/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberCCW0T4
Device Lot Number15511642
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/15/2025
Supplement Date Received by Manufacturer06/16/2025
Initial Report FDA Received Date06/06/2025
Supplement Report FDA Received Date07/10/2025
Date Device Manufactured10/18/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CENTURION VISION SYSTEM ACTIVE SENTRY HANDPIECE.; CLAREON MONARCH IV IOL DELIVERY SYSTEM, INJECTOR.; DUOVISC VISCOELASTIC SYSTEM.; MONARCH II LOADING FORCEPS.; MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D.
Outcome Attributed to Adverse Event Other;
Patient Age66 YR
Patient SexMale
-
-