DEPUY SYNTHES PRODUCTS LLC 11CM ANGLE ATTACHMENT; DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE
|
Back to Search Results |
|
| Catalog Number |
QD11 |
| Medical Device Problem Codes |
Excessive Heating (4030); Difficult or Delayed Separation (4044)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H11: additional narrative: device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the device generating heat, identified during service and evaluation was confirmed.The assignable root cause was determined to be due to component failure from wear.
|
| |
|
Event or Problem Description
|
|
It was noted in the service order that the attachment device was difficult to connect to the cutter device.During service and evaluation, it was determined that the device generated heat.It was further determined that the device had a worn bearing, vibration, component damage, was loose, and made an unexpected noise.It was observed that the cutter could not be inserted.It was further determined that the device failed pretest for visual assessment, thimble set screw, cutter insertion, vibration and temperature.This event did not occur during surgery.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2025.All available information has been disclosed.
|
| |
|
Search Alerts/Recalls
|
|
|