• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS, INC SMART PORT CT; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS, INC SMART PORT CT; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Model Number H787CT96STSA0
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Medical device site infection (4845)
Date of Event 09/21/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The customer's reported complaint description of patient infection cannot be confirmed given the patient centric nature of this serious adverse event (sae).No port device was returned for evaluation since there was no report of port device malfunction, damage or performance issue during use.A device history record (dhr) review of the indicated packaging lot revealed no quality related issues or manufacturing deficiencies at the time of manufacture.Sterilization load release records were reviewed for the packaging lot in question; no issues were observed.The port was implanted into the patient 14 months prior to the patient's infection.There was no report of port device malfunction or performance issue during use.If the manufacturing/packaging/sterilization of the port device in question was to contribute to an infection in the patient it would have likely occurred shortly after the implant date and not 14 months later.Device directions for use (dfu) cautions that each access of the port should be performed using aseptic technique, following the institutions universal precautions.There is no indication from the reported complaint that the manufacturing/packaging/sterilization of the port device could have contributed to the patient's infection.Infection is cautioned in the device dfu as a potential complication of port system use.Labeling review: the directions for use (dfu) that is provided in the smartport device contains the following directions and precautions: contraindications · catheter insertion in the subclavian vein medial to the border of the first rib, an area which is associated with higher rates for pinch-off.Warnings: · the device is to be implanted, used, maintained, and removed in strict accordance with institutional and or centers for disease control (cdc) guidelines or policies.· reuse of single-use devices creates a potential risk of patient or user infections.Contamination of the device may lead to injury, illness or death of the patient.· do not use syringes smaller than 10 ml syringe when accessing the port as system damage can occur.Flushing occluded catheters with small syringes can create excessive pressures within the port system.Absence of a blood return or a poor blood return can be a sign of a potential complication such as occlusion, kinking, breakage, pinch-off syndrome, fibrin formation, thrombosis or malposition.This should be evaluated prior to device usage.· a blood return should be present prior to usage of device for any therapy or testing.· do not attempt to measure the patient's blood pressure on the arm in which a peripheral system is located, since catheter occlusion or other damage to the catheter could occur.· if the patient complains of pain, or there is swelling when the device is flushed or when medication or contrast media is administered, evaluate the device for infiltration, proper needle placement, and potential complications such as occlusion, kinking, breakage, pinch-off syndrome, thrombosis or malposition.Failure to assess these complaints or observations can lead to device failure.Precautions · carefully read and follow all instructions prior to use.· strict aseptic technique is of paramount importance when implanting any device.Potential complications use of an angiodynamics port system involves potential risks normally associated with the insertion or use of any implanted device or indwelling catheter including but not limited to: erosion of vessel and skin implant rejection infection necrosis of scarring of skin over implant area vessel trauma post-operative care the incision site should be monitored for signs of infection, inflammation, hematoma, device rotation or erosion.Routine wound care should be given to these sites.The smart port ct implantable port may be used immediately after verification of catheter placement.Instruct patient to avoid any heavy exertion or strenuous activity during the first few days after surgery.General guidelines · each access of an angiodynamics smart port ct implantable port should be performed using aseptic technique.· follow institutional universal precautions.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
 
Event or Problem Description
On or about (b)(6) 2021, decedent (b)(6) underwent placement of an angiodynamics smartport port catheter product, model number h787ct96stsa0, lot number 5673772.The smartport was implanted by dr.(b)(6), m.D., at (b)(6) hospital - (b)(6), in (b)(6).On or about (b)(6) 2022, decedent presented to (b)(6) hospital - (b)(6), in (b)(6), with symptoms of an infection, subsequently being diagnosed with having an infected port catheter that required removal and in-hospital treatment for the infection.On or about (b)(6) 2022, defendants' defective smartport device was removed from decedent by dr.(b)(6), m.D., at (b)(6) hospital - (b)(6), in (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMART PORT CT
Common Device Name
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
ANGIODYNAMICS, INC
10 glens falls technical park
glens falls NY 12801
Manufacturer (Section G)
ANGIODYNAMICS, INC
10 glens falls technical park
glens falls NY 12801
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
5086587990
MDR Report Key22159562
Report Number1317056-2025-00141
Device Sequence Number12585940
Product Code LJT
UDI-Device Identifier15051684018050
UDI-Public15051684018050
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K062414
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 06/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2022
Device Model NumberH787CT96STSA0
Device Catalogue NumberCT96STSA
Device Lot Number5673772
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/23/2025
Initial Report FDA Received Date06/06/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/07/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention; Other;
Patient SexMale
-
-