• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ROTALINK? PLUS; CATHETER, CORONARY, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ROTALINK? PLUS; CATHETER, CORONARY, ATHERECTOMY Back to Search Results
Model Number H749236310020
Medical Device Problem Codes Fluid/Blood Leak (1250); Noise, Audible (3273)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/05/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a catheter saline leak occurred.The 95% stenosed target lesion was located in the moderately tortuous and severely calcified left anterior descending artery (lad).A 1.25mm rotablator rotalink plus was selected for use.Before the draw test, during the rotational check, the test could not be completed.When rotation was switched on, a sudden, different noise occurred, and then the machine automatically went into dyna mode and shut off.A saline leak was then observed, but the location of the leak was unknown.The procedure was completed with another device of the same brand.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTALINK? PLUS
Common Device Name
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
model farm road
cork T12 Y K88
EI   T12 YK88
Manufacturer Contact
mark mischke
4100 hamline ave n
arden hills, MN 55112
7634941632
MDR Report Key22160061
Report Number2124215-2025-36112
Device Sequence Number7422016
Product Code MCX
UDI-Device Identifier08714729316411
UDI-Public(01)08714729316411(17)261112(10)35480313
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
P900056
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberH749236310020
Device Catalogue NumberH749236310020
Device Lot Number0035480313
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/12/2025
Supplement Date Received by Manufacturer06/30/2026
Initial Report FDA Received Date06/06/2025
Supplement Report FDA Received Date07/22/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/12/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age77 YR
Patient SexFemale
-
-