| Catalog Number |
SRM-210 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported "rupture ".This record is for the right side.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: b5, d4, d6b, h4, h6.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side rupture.Device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: d9 h3 h6.Laboratory analysis summary the device related to the reported events rupture was received on jul 22, 2025,with lot number 3091244.Based on the product analysis performed, the assessments of the complaint codes are: ¿ rupture: observed an opening assessed as fold flaw opening.As per the investigation procedure, crease and non penetrating nick wear observed and none of the observations are found to be potentially related to the manufacturing process, no further actions are required.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "rupture".This record is for the right side.Device has been explanted and replaced.
|
| |
|