• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. BD SYRINGE 50CC LL TIP CONVENIENCE PAK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. BD SYRINGE 50CC LL TIP CONVENIENCE PAK Back to Search Results
Catalog Number 309680
Medical Device Problem Code Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/15/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
 
Event or Problem Description
It was reported that bd syringe 50cc ll tip convenience pak had foreign matter.It was reported by the customer that unknown particles were observed in two syringes.Verbatim: rcc received a complaint via email.Email(s) attached.Incoming inspection of several lots of 309680 revealed two defects.A syringe from lot 4355027 contained a blue fiber on the rubber stopper within the barrel of the assembly.A syringe from lot 4355030 showed black particulates on the plunger just above the rubber stopper.
 
Additional Manufacturer Narrative
Investigation results: our quality engineer inspected the 2 photos submitted for evaluation.The reported issue of foreign matter was confirmed upon inspection of the samples.Analysis of the samples showed that one of the photos show a syringe plunger rod with dark colored specks and the other photo shows a syringe rubber stopper with a red colored hand drawn circle that is highlighting what appears to be a foreign matter adhered to the rubber stopper.Bd is not able to provide a probable root cause without the evaluation of a physical sample.A review of our risk management documentation was performed and indicates that the potential risk of the reported event was assessed appropriately.Production records were reviewed, and this batch meets our manufacturing specification requirements.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
No additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD SYRINGE 50CC LL TIP CONVENIENCE PAK
Common Device Name
SYRINGE
Manufacturer (Section D)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX 
Manufacturer (Section G)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX  
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key22167481
Report Number9610847-2025-00176
Device Sequence Number9383041
Product Code FMF
UDI-Device Identifier00382903096800
UDI-Public(01)00382903096800(11)250205(17)291231(10)4355027
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K110771
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number309680
Device Lot Number4355027
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/19/2025
Supplement Date Received by Manufacturer07/29/2025
Initial Report FDA Received Date06/09/2025
Supplement Report FDA Received Date07/31/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/05/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-