| Model Number |
LF5644 |
| Medical Device Problem Codes |
Difficult to Open or Close (2921); Intermittent Energy Output (4025)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/12/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a procedure involving the sealing devices, about 1 hour into the surgery, the jaws of the two devices got locked close or at the uterine vessels and could not open for sealing and excising the fallopian tube.The devices were difficult or impossible to open and got stuck on tissue, with the jaws stuck closed.The devices slipped off the tissue and were able to be removed with the jaws closed.The devices did not have to be cleaned as there had been no build-up of eschar.The knife blades were not extended and did not protrude beyond the edge of the jaws.This incident did not cause the procedure to be extended by more than 30 minutes.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
D10 concomitant product: lf5644, lig dissector lf5644 sealer div 44cm (lot#40580133x) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during a procedure involving the ligasure vessel sealing devices, about 1 hour into the surgery, the jaws of the two devices got locked close or at the uterine vessels and could not open for sealing and excising the fallopian tube.The devices were difficult or impossible to open and got stuck on tissue, with the jaws stuck closed.The devices slipped off the tissue and were able to be removed with the jaws closed.The devices did not have to be cleaned as there had been no build-up of eschar.The knife blades were not extended and did not protrude beyond the edge of the jaws.This incident did not cause the procedure to be extended by more than 30 minutes.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5, d3 (mfr name and address), d4 (udi#), g3, h3, h6.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that that the jaws was difficult to open and the device stopped working.The reported issues could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|