Model Number BREVDRV
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/12/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported on (b)(6) 2025, that during a brevera procedure, the device driver would not fire after the patient was already in compression and the needle had already been inserted.It is reported that the customer attempted to troubleshoot the issue with the assistance of hologic technical support via the telephone; however, it did not resolve the issue.As a result, the procedure could not be completed, and the patient was rescheduled.A field engineer was dispatched to assess the device, and found the device was indicating that the device driver was not recognized.The field engineer indicates that a review of system logs showed several errors associated with needle removal and the array not having communication with the system.It is reported that the field engineer performed system calibrations, which allowed the device driver to be recognized.The field engineer confirmed that the system is now ready for use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BREVERA BREAST BIOPSY SYSTEM
Common Device Name
INSTRUMENT, BIOPSY
Manufacturer (Section D)
HOLOGIC, INC
250 campus drive
marlborough MA 01752
Manufacturer (Section G)
HOLOGIC, INC
250 campus drive
marlborough MA 01752
Manufacturer Contact
dara smithers
600 technology drive
newark, DE 19702
MDR Report Key22168126
Report Number1222780-2025-00306
Device Sequence Number7818342
Product Code KNW
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K163052
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberBREVDRV
Device Catalogue NumberBREVDRV
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/12/2025
Supplement Date Received by Manufacturer07/08/2026
Initial Report FDA Received Date06/09/2025
Supplement Report FDA Received Date07/30/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/20/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-