| Catalog Number |
ECR45B |
| Medical Device Problem Codes |
Failure to Form Staple (2579); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Codes |
Hemorrhage/Blood Loss/Bleeding (1888); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/28/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an unknown procedure, it was observed that a few staples formed with irregular shapes and anastomotic site was oozing after firing.Changed to another one to continue the procedure but the same problem happened again.To stop oozing with compression hemostasis.There was no patient consequence nor surgical delay reported.No a.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 06/09/2025.D4: batch #unknown.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device ecr45b with lot number 307c36; and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Other remarks.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4) date sent: 06/27/2025 d4: batch #229c23.Corrected data: h4.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one ecr45b reload was received with no apparent damage, fully fired and open package.No functional test could be performed due to the condition of the reload.The event described could not be confirmed as the reload was received fully fired.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.Device history review a manufacturing record evaluation was performed for the finished device batch number 229c23, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|