| Catalog Number |
NSLX120L |
| Medical Device Problem Codes |
Difficult to Open or Close (2921); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an unknown procedure the device locked up.The jaw would not open, and it was tearing the tissue on the removal.They where able to open and replace it with a new one.No adverse impact to patient/user was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 6/9/2025 b3: event year only reported: 2025.D4 batch #: z70096.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformance were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4) date sent: 7/21/2025 d4 batch #: z70096.Investigation summary the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the device was received with no apparent damage.The device was tested on the generator and passed all functional testing.The energy output delivered from the device was verified.All tones were heard during functional testing.The device was tested with the test media and no anomalies were found.No issues were noted with the opening and closing of the jaws.There were no anomalies noted with the functionality of the device.As part of the quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as the device performed without any difficulties noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.A manufacturing record evaluation was performed for the finished device batch z70096, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|