• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 DUAL MOBILITY LINER 40MM D; PROSTHESIS, HIPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 DUAL MOBILITY LINER 40MM D; PROSTHESIS, HIPS Back to Search Results
Model Number N/A
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 05/15/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D10: ep-200146 act artic e1 hip brg 28x40mm (b)(6), 110010243 g7 osseoti 3 hole shell 50mm d (b)(6), unknown ceramic head.G2: foreign: australia.Customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported the patient underwent a hip arthroplasty and was implanted with zimmer biomet products.Subsequently, the patient was revised nearly one (1) year post implantation due to dislocation.The liner was explanted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected.Information.Updated: d4, g3, g6, h2, h3, h4, h6, h10, h1.Visual examination of the provided pictures identified the bearing and head implants covered in bio-debris.No further evaluation could be made from the pictures provided.Review of the device history records identified no deviations or anomalies during manufacturing.An undated radiograph was provided and not submitted to mmi due to the inability to correlate the image with the allegation.A definitive root cause cannot be determined.Complaint cannot be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor trends.
 
Event or Problem Description
No additional event information to report at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 DUAL MOBILITY LINER 40MM D
Common Device Name
PROSTHESIS, HIPS
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key22168787
Report Number0001822565-2025-01741
Device Sequence Number21134443
Product Code LPH
UDI-Device Identifier00880304703506
UDI-Public(01)00880304703506(17)330928(10)66310817
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K150522
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberN/A
Device Catalogue Number110024462
Device Lot Number66310817
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/15/2025
Supplement Date Received by Manufacturer09/04/2025
Initial Report FDA Received Date06/09/2025
Supplement Report FDA Received Date09/16/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/28/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11 NARRATIVE.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
Patient Weight58 KG
-
-