| Model Number |
7379 |
| Medical Device Problem Code |
Missing Test Results (3267)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/14/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The patient/family (a1) was the initial reporter (e1), so personal information was not entered.No information was captured in section a4 as the customer's weight was not provided.The contour® next meter does not have an expiration date.Therefore, no information was captured in section d4 (expiration date).The warranty period of the meter is 5 years from the date of the original purchase.
|
| |
|
Event or Problem Description
|
|
The customer reported that her blood glucose readings were not being stored in the memory of the contour® next meter.There was no allegation of an adverse event.The customer was advised to return the device for evaluation.A replacement meter kit was sent to the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
Despite three follow-up attempts, the customer did not return the device for evaluation.Therefore, a device history record was reviewed for the suspected contour® next meter, serial # (b)(6), and no manufacturing anomalies were found.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer returned the suspected contour® next meter (serial # (b)(6)) for evaluation.No test strips were returned.An in-house testing was performed with the returned meter and in-house contour® next test strips (lot # 3mneg02a) using blood spiked with glucose at 135 mg/dl, as determined by ysi instrument in five replicates.All tests recovered within fit-for-use limits.The blood glucose results obtained with the in-house testing were properly stored in the meter's memory.
|
| |
|