It was reported that on (b)(6) 2025, during a brevera procedure, the user experienced an imaging system error, and the x-ray went down.It is reported that the user was unable to clear the error, restarted the system and completed another qc on the needle; however, received another imaging system error as well as a filter error.The user reports that the system would not allow the procedure to move forward and the patient was rescheduled after the needle had already been inserted.A field engineer was dispatched to assess the device but was unable to reproduce the error.The field engineer utilized a test needle and was able to take 12 consecutive exposures, and the system passed the needle test.The field engineer concluded that the system had a disconnect with the corlumina and a shutdown with power off for more than a few minutes made it so that the system was able to re-establish a link once power was restored.No further information is available.
|