| Catalog Number |
448010 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/16/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported while using the bd phoenix¿ ap contamination occurred due to pipette tips colliding with one another.No health impact or consequence reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: the complaint on dispense tubing contamination was reported in phoenix ap instrument.Field service engineer (fse) was dispatched on site.Fse replaced the dispense tubing and issue resolved.Instrument is fully operational.This is a confirmed failure of a bd product.The root cause of the failure is not known.Device history record review for the instrument (b)(6) is not required for this complaint as this complaint does not allege an early life failure or failure at installation and the configuration has changed since release from manufacturing due to service repairs/pms.Service history review was completed and revealed no previous complaints related to this failure mode.Samples were not received by quality for investigation and thus, returned material investigation could not occur.If samples are received at a later date, the complaint may be reopened.Bd quality will continue to closely monitor for trends associated with this complaint.
|
| |
|
Event or Problem Description
|
|
It was reported while using the bd phoenix¿ ap contamination occurred due to pipette tips colliding with one another.No health impact or consequence reported.
|
| |
|