COCHLEAR BONE ANCHORED SOLUTIONS AB BI300 IMPLANT 4 MM; COCHLEAR BAHA CONNECT SYSTEM Back to Search Results
Model Number 92129
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code Implant Pain (4561)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reported adverse event is associated with a returned device; however, the provided clinical information was reviewed by the manufacturer and no specific device analysis is deemed necessary at this time.Previous product examinations have not showed any relationship between a product geometrical deviation and the reported clinical complication.Additionally, there are no indications that a product failure has contributed to the reported issue.Transient pain is a known medical complication where most cases can be resolved with clinical case management.Known reasons for cause of pain are abnormal skin sensitivity, earlier trauma to the implanted area, lack of adequate bone quality or other generalized diseases.It might also be due to a loose implant magnet, causing pain around the implanted area when using the baha sound processor.
 
Event or Problem Description
Per the clinic, the patient experienced chronic pain from the implant (specific date not reported).Subsequently, the patient was placed under general anesthesia on (b)(6) 2025, where implant was explanted.The abutment was also removed during the same surgery and the site was sutured off.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BI300 IMPLANT 4 MM
Common Device Name
COCHLEAR BAHA CONNECT SYSTEM
Manufacturer (Section D)
COCHLEAR BONE ANCHORED SOLUTIONS AB
konstruktionsvägen 14
mölnlycke, 435 3 3
SW  435 33
Manufacturer (Section G)
COCHLEAR BONE ANCHORED SOLUTIONS AB
konstruktionsvägen 14
mölnlycke, 435 3 3
SW   435 33
Manufacturer Contact
thanaletchumi manogaran
unit ug-1, vertical podium
no. 8 jalan kerinchi,
kuala lumpur, wilayah persekutuan 59200
MY   59200
MDR Report Key22173139
Report Number6000034-2025-02084
Device Sequence Number7957683
Product Code LXB
UDI-Device Identifier09321502019545
UDI-Public(01)09321502019545(10)COH1233460(17)230516
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K100360
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date05/16/2023
Device Model Number92129
Device Catalogue Number92129
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/15/2025
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/15/2025
Initial Report FDA Received Date06/09/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-