• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® MOLDING & OCCLUSION BALLOON CATHETER; CATHETER, INTRAVASCULAR OCCLUDING, TEMPORARY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® MOLDING & OCCLUSION BALLOON CATHETER; CATHETER, INTRAVASCULAR OCCLUDING, TEMPORARY Back to Search Results
Catalog Number MOB37
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Vascular Dissection (3160)
Date of Event 05/15/2025
Type of Reportable Event Serious Injury
Event or Problem Description
The following information was reported to gore: on (b)(6) 2025, this patient underwent endovascular treatment for an abdominal aortic aneurysm using gore® excluder®aaa endoprostheses and gore® molding & occlusion balloon catheter as an accessory.After the aortic extender was deployed, the physician inflated the mob balloon inside.Subsequently, an angiography revealed retrograde dissection extending from the proximal end of the aortic extender to below the right renal artery.As the dissection was localized, the procedure was completed with blood pressure management and follow-up observation.The physician¿s comment: ¿the cause is believed to be an overly forceful touch-up.¿.
 
Additional Manufacturer Narrative
H6: a review of the manufacturing records for the device verified that the lot met all pre-release specifications.H6: code d12: according to the gore® molding & occlusion balloon catheter instructions for use, adverse events that may occur and / or require intervention or additional intraoperative procedure time include but are not limited to: trauma to the vessel wall, including spasm, dissection, perforation, or rupture.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may have not been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® MOLDING & OCCLUSION BALLOON CATHETER
Common Device Name
CATHETER, INTRAVASCULAR OCCLUDING, TEMPORARY
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL PHOENIX 1 B/P
32360 n. north valley parkway
phoenix AZ 85085
Manufacturer Contact
eimi ueda
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key22173164
Report Number3007284313-2025-04022
Device Sequence Number9218542
Product Code MJN
UDI-Device Identifier00733132639489
UDI-Public(01)00733132639489(17)271030
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K172567
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberMOB37
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/15/2025
Initial Report FDA Received Date06/09/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/30/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age86 YR
Patient SexFemale
-
-