| Model Number |
H74939171080210 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/27/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D2b - pro code (product code): fge, lit.G4 - premarket / 510(k) #: k141521, k141597.
|
| |
|
Event or Problem Description
|
|
It was reported that a balloon rupture occurred.The 99% stenosed target lesion was located in the mildly tortuous and severely calcified iliac artery.An 8.0 x 20, 135cm mustang balloon catheter was advanced for dilation.During the procedure, the balloon ruptured upon second inflation at the nominal burst pressure for 10 seconds.The device was removed without any problem, and the procedure was completed with another of the same device.There were no patient complications as a result of this event.
|
| |
|
Search Alerts/Recalls
|