| Catalog Number |
CDH21A |
| Medical Device Problem Codes |
Failure to Form Staple (2579); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Codes |
Hemorrhage/Blood Loss/Bleeding (1888); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/24/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4) date sent 6/10/2025 d4 batch # unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained: 1.Could you please provide more details about the type of oozing? 2.How much blood was lost (ml)? unknown.3.Did the patient require a transfusion? no.4.Could you please clarify if there were any patient consequences? no.5.Could you please clarify if there were any change in the post-operative care of the patient as a result of the event? no.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this lot.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during an esophageal surgery.After fired, noted that a few staples formed with irregular shapes and anastomotic site was oozing.To stop oozing with compression hemostasis.There was no patient consequence reported, no additional information could be provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent 7/18/2025.The device upon which this medwatch is based has been received, however, the evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent 7/21/2025.D4 batch #148d71.Corrected data d4: udi#.Investigation summary : the product was returned for thorough evaluation, which included visual inspections and functional testing of the returned device.Visual analysis of the returned sample determined that the cdh21a device arrived with no apparent damage along with the stapler retainer not placed.The breakaway washer uncut and indented and there were no staples present.The device was reloaded with staples and tested for functionality with the returned washer; the device formed all the staples, as well as completely cut the test media and the breakaway washer without incident.The staple line was complete, and the staples meet the staple form release criteria.The event reported was confirmed and it is related to improper use of the device.The condition of the washer indicates that the device had not been fired through a full firing stroke or possibly that the orange indicator was not fully into the safe green firing range.Before firing the device the orange indicator should be fully within the green range of the gap setting scale and the safety should not be released prior to firing.Also, if the firing sequence is not fully completed (the firing handle reaches its stopping point, and the firing trigger is parallel to the instrument handle) staples could be partially deployed without cutting the washer.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 148d71, and no non-conformances were identified.
|
| |
|