| Model Number |
GIA6038S |
| Medical Device Problem Codes |
Failure to Fire (2610); Component or Accessory Incompatibility (2897); Human-Device Interface Problem (2949)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/15/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a procedure, the device was difficult to load on the second loading unit.When it was time to staple, the device was unable to fire and was difficult to remove the reload in place.The device was replaced to resolve the issue.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
D10 concomitant product: unknown gia, unknown gia product (lot#unknown).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d3 (mfr name and street 1), d4(udi), d9, g3, h3, h6 h3 evaluation.Summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection of the instrument noted no abnormalities.The single use loading unit (sulu) was partially applied.The interlock was engaged.Microscopic inspection of the knife blade displayed damage.Functional testing noted that the sulu was reset and loaded into the instrument.The sulu unloaded and loaded repeatedly without difficulty.The sulu and instrument were applied to test media with acceptable results.The firing knob of the instrument retracted and advanced properly without difficulty.The knife cut completely and cleanly and the staple line was properly formed.It was reported that the instrument did not fire.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur if the loading unit and knife blade was applied over an obstruction or thick tissue during clinical application.It was also reported that the instrument was difficult to load and difficult to unload.The reported issue could not be confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: the instrument should not be used on tissue such as liver or spleen where compressibility is such that closure of the instrument would be destructive.Tissue thickness should be carefully evaluated before firing any stapler.If the tissue can not comfortably compress to the specified minimum requirement, or compresses to less than this requirement, the tissue may be too thick or too thin for the selected staple size.Do not rotate the firing knob during firing.Rotating the knob during firing may cause damage and possible instrument malfunction.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|