| Catalog Number |
0250070453 |
| Medical Device Problem Codes |
Break (1069); Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/18/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that the outer plastic sheath broke inside the patient's abdomen during surgery.The broken piece was successfully retrieved.
|
| |
|
Additional Manufacturer Narrative
|
|
The sales representative confirmed that the device will not be returned.This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: broke inside of patient (successfully removed from patient) probable root cause: material/design error, excessive user force, severe shipping conditions, disposable tip rubbing against product, user error.The reported failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that the outer plastic sheath broke inside the patient's abdomen during surgery.The broken piece was successfully retrieved.
|
| |
|