MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGACLIP*MCA MED SHORT APPLIER; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGACLIP*MCA MED SHORT APPLIER; CLIP, IMPLANTABLE Back to Search Results
Catalog Number MSM20
Medical Device Problem Code Failure to Form Staple (2579)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/29/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Date sent 6/10/2025 d4 batch # unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this lot.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: did device jammed (not fire clips)? did device not feed clips? did device drop or eject clips? did device sideways feed clips? did device fire malformed clips? did device fire scissored clips? if other please specify this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during an unknown procedure there were loose clips when applying.Some falling out¿.Very generic feedback and no specific details.No surgical issues as a result.There was no surgical delay.There were no patient consequences.
 
Additional Manufacturer Narrative
(b)(4).Date sent 6/27/2025.The device upon which this medwatch is based has been received, however, the evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.
 
Additional Manufacturer Narrative
(b)(4) date sent 7/2/2025 d4 batch #y7028z.Investigation summary the product was returned for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that the msm20 devices was returned inside it's packaging unopened with a clip in jaws.Upon visual inspection, it was observed that the tyvek and the blister from the packaging were damaged; it was noted to have a hole in the tyvek, the hole was noted to be from the outside in.In an attempt to replicate the reported incident, the device was tested for functionality.Upon testing, the packaging was opened and then the device was cycled and it fed and formed the remaining twenty(20) clips as intended.As part of ethicon¿s quality process all devices are manufactured, inspected, and released to approved specifications.Due to the damages found on the packaging, a possible cause for these conditions is due to improper handling during transit or storage; it appears that the package hit a pointy surface and this caused the reported event.Although no conclusion could be reach on the cause of the reported event, the instructions for use do contain the following caution: ensure that a clip is in the jaws.Position the jaws so that the clip completely surrounds the vessel to be ligated.Fully squeeze the handles together until they stop.As the handles are squeezed, the clip closes and occludes the vessel or tubular structure.Fully release the handles to return the instrument to its original open position.The next clip is automatically advanced.Inspect the jaws after each clip application to ensure the presence of a new clip before applying the next clip.Repeat steps as necessary to continue ligation of tissue.For optimal performance, periodically clean the instrument during the procedure by submerging the jaw and distal end of the applier shaft in saline.The event described could not be confirmed as the device performed without any difficulties noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.A manufacturing record evaluation was performed for the finished device batch y7028z number, and no non-conformances were identified.
 
Additional Manufacturer Narrative
(b)(4) date sent: 7/7/2025.Additional information was requested, and the following was obtained: how did device not work? did device jammed (not fire clips)? did device not feed clips? did device drop or eject clips? did device sideways feed clips? did device fire malformed clips? did device fire scissored clips? if other please specify when information requested, the staff had nothing else to add.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
3035526892
MDR Report Key22177863
Report Number3005075853-2025-04388
Device Sequence Number15802972
Product Code GDO
UDI-Device Identifier10705036002505
UDI-Public(01)10705036002505(17)290630(10)182D96
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K820837
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 07/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberMSM20
Device Lot Number182D96
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/06/2025
Supplement Date Received by Manufacturer06/26/2025
07/01/2025
07/02/2025
Initial Report FDA Received Date06/10/2025
Supplement Report FDA Received Date06/27/2025
07/02/2025
07/07/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/02/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-