Catalog Number 320-15-01
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Inadequate Osseointegration (2646); Metal Related Pathology (4530)
Date of Event 03/12/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
D10: concomitants: (b)(6), 320-06-38 - glenosphere 38mm, (b)(6), 320-15-05 - eq rev locking screw, (b)(6), 320-20-00 - eq reverse torque defining screw kit, (b)(6), 320-06-38 - glenosphere 38mm, (b)(6), 320-15-05 - eq rev locking screw, (b)(6), 320-20-00 - eq reverse torque defining screw kit, (b)(6), 320-20-26 - eq rev compress screw lck cap kit, 4.5 x 26mm, (b)(6), 320-20-30 - eq rev compress screw lck cap kit, 4.5 x 30mm, (b)(6), 320-38-13 - 145-deg pe 38mm const hum liner +2.5, (b)(6), 321-20-00 - equinoxe reverse shoulder drill kit, (b)(6), 320-10-00 - equinoxe reverse tray adapter plate tray +0, (b)(6), 300-01-11 - equinoxe, humeral stem primary, press fit 11mm.Should additional relevant information be obtained, a follow-up mdr will be submitted accordingly.
 
Event or Problem Description
It was reported a 68 yo male patient, who had a left shoulder implanted, underwent a revision procedure.It appeared the humeral liner was dissociated from the humeral plate.The surgeon believed that the poly had been dissociated for quite some time per the x-rays.The patient had significant metallosis throughout the joint.Both the humeral stem and glenoid baseplate were loose.They were removed and an antibiotic spacer was implanted.Patient will be revised at a later date with a custom glenoid.The patient was revised to a competitor¿s devices.There were no device breakages or surgical delays during the procedure.The patient was last known to be in stable condition following the event.X-rays were provided.The explanted devices are not available for return due to hospital policy.Device images were provided.No further information.This is 1 of 5 reports for this event.This is the 2nd of 2 events for this patient.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated/corrected: d1, h6.Mdr section codes updated/corrected: b, c, d, g.The revision reported was likely the result of disassembly between the humeral liner and humeral tray leading to subsequent metal-on-metal articulation between unintended components.Additional reasons for the reported revision may be inclusion of the polyethylene in the packaging recall.However, the reported loosening of the glenoid baseplate and humeral stem and observed metallosis in the joint cannot be confirmed based on the provided information.Potential contributions from patient conditions, an unreported traumatic event, incomplete seating of the humeral liner during implantation, forceful contact between the liner and glenoid bone (scapular notching), or a combination of the above to the disassembly of the humeral liner cannot be confirmed from the reported information.Should additional relevant information be obtained, a follow-up mdr will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQ REV GLENOID PLATE
Common Device Name
PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66 court
gainesville, FL 32653
3525541928
MDR Report Key22178375
Report Number1038671-2025-02267
Device Sequence Number8678953
Product Code KWT
UDI-Device Identifier10885862086488
UDI-Public10885862086488
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K063569
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number320-15-01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/23/2025
Supplement Date Received by Manufacturer10/10/2025
Initial Report FDA Received Date06/10/2025
Supplement Report FDA Received Date10/15/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/21/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
Patient SexMale
-
-