| Model Number |
FX5010SWC |
| Medical Device Problem Code |
Failure to Osseointegrate (1863)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Unspecified Infection (1930); Inadequate Osseointegration (2646)
|
| Date of Event |
04/30/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The doctor reported the implant was removed due to osseointegration failure, 3 months after implant placement.Bone quality around the area was type iv, and oral hygiene around the implant was moderate.Local infection and peri-implantitis were observed during the implant removal surgery.Bone augmentation material was used in implant placement surgery.The patient will need another surgical intervention.The doctor did a socket presentation procedure after the implant came out.
|
| |
|
Additional Manufacturer Narrative
|
|
Please see attached summary memo.
|
| |
|