| Catalog Number |
SSL-350 |
| Medical Device Problem Codes |
Device Appears to Trigger Rejection (1524); Material Rupture (1546)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
01/08/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iii, rupture.A review of the device history record has been completed.No deviations or non-conformances noted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "ruptured implant with capsular contracture baker grade iii" confirmed via ultrasound.This record is for the right side.The device remains implanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side rupture and capsular contracture, baker grade iii.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: b5; b7; d6b; h6.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: b.5, b.6, h.6.
|
| |
|
Event or Problem Description
|
|
Previous medwatch submission noted rupture.Upon further information, allergan has determined that the event of rupture did not occur.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "ruptured implant with capsular contracture baker grade iii" confirmed via ultrasound.Healthcare professional later reported "intact".This record is for the right side.The device was explanted, replaced and returned.
|
| |
|
Additional Manufacturer Narrative
|
|
Lab analysis: per the investigation procedure, the device is analyzed through visual inspection microscopic inspection if openings are observed and a weight verification.Per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required: ¿ capsular contracture: unable to observe through visual inspection as it is a physiological phenomenon.None of the other observations performed during the device analysis deformation and creases are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
|
| |
|