(b)(4).Date sent: 06/10/2025 d4: batch # 489c84 investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one ecr60w reload was received partially fired 1/3, with the pan detached from the reload.In addition, some drivers missing, making the reload non-functional.No functional test was performed due to the condition of the reload.The partially fired is consistent with an incomplete or interrupted cycle.It is possible that handling by the customer could dislodge the pan.Dislodgement as a result of the removal of the reload from the device is the most likely scenario.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device ecr60w with lot number 519c15; and no non-conformances were identified.A manufacturing record evaluation was performed for the finished device batch number 489c84, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|