| Model Number |
102-9800 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/04/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device analysis of the returned driver revealed that the tips of the driver were completely sheared off, as confirmed during visual inspection.As such, this damage was sufficient to prevent functional testing in the laboratory.This confirmed damage to the driver was likely an unintended result of excessive force while using the device.A review of the instructions for use (ifu) was performed, it states to avoid application of excessive stress on instrumentation and do not force deployment or implant breakage or damage to bony structures may result.
|
| |
|
Event or Problem Description
|
|
It was reported that during the superion indirect decompression system (ids) implant procedure, the prongs of the driver broke during deployment of the implant.The physician assessed that this damage was due to the patients narrow anatomy which caused some resistance.All broken pieces of the driver were removed from the patient successfully however, the procedure was aborted due to the patients difficult anatomy.
|
| |
|
Event or Problem Description
|
|
It was reported that during the superion indirect decompression system (ids) implant procedure, the prongs of the driver broke during deployment of the implant.The physician assessed that this damage was due to the patients narrow anatomy which caused some resistance.All broken pieces of the driver were removed from the patient successfully however, the procedure was aborted due to the patients difficult anatomy.
|
| |
|
Additional Manufacturer Narrative
|
|
A retroactive review of events within the quality system identified this record as requiring remediation, including reporting and submission of an initial emdr per applicable criteria.Boston scientific has opened a corrective and preventive action (capa 8647) investigation to determine the root cause for why an initial emdr was not previously submitted to the fda in accordance with boston scientific established mdr reporting process and fda adverse event reporting requirements.Information gained through these investigation efforts will be utilized to determine root cause and implement appropriate preventive action(s), as necessary.Fields h7 and h9 were updated.Device analysis of the returned driver revealed that the tips of the driver were completely sheared off, as confirmed during visual inspection.As such, this damage was sufficient to prevent functional testing in the laboratory.This confirmed damage to the driver was likely an unintended result of excessive force while using the device.A review of the instructions for use (ifu) was performed, it states to avoid application of excessive stress on instrumentation and do not force deployment or implant breakage or damage to bony structures may result.
|
| |
|
Additional Manufacturer Narrative
|
|
A retroactive review of events within the quality system identified this record as requiring remediation, including reporting and submission of an initial emdr per applicable criteria.Boston scientific has opened a corrective and preventive action (capa 8647) investigation to determine the root cause for why an initial emdr was not previously submitted to the fda in accordance with boston scientific established mdr reporting process and fda adverse event reporting requirements.Information gained through these investigation efforts will be utilized to determine root cause and implement appropriate preventive action(s), as necessary.Fields h7 and h9 were updated.Device analysis of the returned driver revealed that the end of the driver was damaged, and the tips of the driver were completely sheared off, as confirmed during visual inspection.As such, this damage was sufficient to prevent functional testing in the laboratory.This confirmed damage to the driver was likely an unintended result of excessive force while using the device.A review of the instructions for use (ifu) was performed, it states to avoid application of excessive stress on instrumentation and do not force deployment or implant breakage or damage to bony structures may result.It also states to carefully inspect all instruments prior to and after use.Do not use an instrument that is visibly damaged.If an instrument appears damaged after use confirm that no broken fragments are retained in the patient.
|
| |
|
Event or Problem Description
|
|
It was reported that during the superion indirect decompression system (ids) implant procedure, the prongs of the driver broke during deployment of the implant.The physician assessed that this damage was due to the patient's narrow anatomy which caused some resistance.All broken pieces of the driver were removed from the patient successfully; however, the procedure was aborted due to the patient's difficult anatomy.
|
| |
|