| Model Number |
102-9800 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/20/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the driver broke during the superion indirect decompression system (ids) implant procedure.All broken fragments were successfully removed.There were no signs of excessive force, sagittal movement, or gear-shifting during the procedure, and no bone or osteophytes were present to cause resistance.The patient did not experience any reported complications.The damaged product was replaced and sent for analysis, and the procedure was completed without further issues.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis of the returned driver revealed that the end of the driver was damaged and completely sheared off.The damage to the driver was sufficient to prevent functional testing.Scanning electron microscopy (sem) analysis, material composition, and hardness testing revealed the device anomaly is associated to the presence of excessive hydrogen in the instrument material, likely due to a manufacturing deficiency.A review of the instructions for use (ifu) was performed, it states to avoid application of excessive stress on instrumentation and do not force deployment or implant breakage or damage to bony structures may result.
|
| |
|
Additional Manufacturer Narrative
|
|
A retroactive review of events within the quality system identified this record as requiring remediation, including reporting and submission of an initial emdr per applicable criteria.Boston scientific has opened a corrective and preventive action (capa 8647) investigation to determine the root cause for why an initial emdr was not previously submitted to the fda in accordance with boston scientific established mdr reporting process and fda adverse event reporting requirements.Information gained through these investigation efforts will be utilized to determine root cause and implement appropriate preventive action(s), as necessary.Fields h7 and h9 were updated.An urgent medical device correction (97003015-fa) was delivered to physicians in may 2023 communicating the potential for superion indirect decompression system (ids) driver instrument tip breaks to occur with use of excessive force during the implant procedure.Driver tip damage may be readily detected via visual, audible, and/or tactile signals.Analysis of the returned driver revealed that the end of the driver was damaged and completely sheared off.The damage to the driver was sufficient to prevent functional testing.Scanning electron microscopy (sem) analysis, material composition, and hardness testing revealed the device anomaly is associated to the presence of excessive hydrogen in the instrument material, likely due to a manufacturing deficiency.A review of the instructions for use (ifu) was performed, it states to avoid application of excessive stress on instrumentation and do not force deployment or implant breakage or damage to bony structures may result.
|
| |
|
Event or Problem Description
|
|
It was reported that the driver broke during the superion indirect decompression system (ids) implant procedure.All broken fragments were successfully removed from the patient and the procedure was completed successfully with a new driver.There were no reported patient complications.
|
| |
|
Event or Problem Description
|
|
It was reported that the driver broke during the superion indirect decompression system (ids) implant procedure.All broken fragments were successfully removed from the patient and the procedure was completed successfully with a new driver.There were no reported patient complications.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction to the follow-up 1 mdr in block h6 and h11.A retroactive review of events within the quality system identified this record as requiring remediation, including reporting and submission of an initial emdr per applicable criteria.Boston scientific has opened a corrective and preventive action (capa 8647) investigation to determine the root cause for why an initial emdr was not previously submitted to the fda in accordance with boston scientific established mdr reporting process and fda adverse event reporting requirements.Information gained through these investigation efforts will be utilized to determine root cause and implement appropriate preventive action(s), as necessary.An urgent medical device correction (97003015-fa) was delivered to physicians in (b)(6)communicating the potential for superion indirect decompression system (ids) driver instrument tip breaks to occur with use of excessive force during the implant procedure.Driver tip damage may be readily detected via visual, audible, and/or tactile signals.Analysis of the returned driver revealed that the end of the driver was damaged and completely sheared off.The damage to the driver was sufficient to prevent functional testing.Scanning electron microscopy (sem) analysis, material composition, and hardness testing revealed the device anomaly is associated to the presence of excessive hydrogen in the instrument material, likely due to a manufacturing deficiency.Additionally, a labeling review was performed and revealed no anomalies.The instructions for use (ifu) states to avoid application of excessive stress on instrumentation and do not force deployment or implant breakage or damage to bony structures may result.Furthermore, it indicates instruments should be carefully inspected prior to and after use; and not to use an instrument that is visibly damaged.If an instrument appears damaged after use it should be confirmed that no broken fragments are retained in the patient.
|
| |
|
Search Alerts/Recalls
|