| Model Number |
102-9800 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/29/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Analysis of the returned driver revealed that the end of the driver was damaged and both tips of the driver were completely sheared off.The damage to the driver was sufficient to prevent functional testing.The damage to the driver suggests that excessive force was used during the procedure.A review of the instructions for use (ifu) was performed, it states to avoid application of excessive stress on instrumentation and do not force deployment or implant breakage or damage to bony structures may result.
|
| |
|
Event or Problem Description
|
|
It was reported that during the superion indirect decompression system (ids) implant procedure, the driver tip broke.All broken fragments were successfully removed.The broken driver was replaced with a new one and the procedure was completed without further issues.
|
| |
|
Event or Problem Description
|
|
It was reported that during the superion indirect decompression system (ids) implant procedure, the driver tip broke.All broken fragments were successfully removed.The broken driver was replaced with a new one and the procedure was completed without further issues.
|
| |
|
Additional Manufacturer Narrative
|
|
A retroactive review of events within the quality system identified this record as requiring remediation, including reporting and submission of an initial emdr per applicable criteria.Boston scientific has opened a corrective and preventive action (capa 8647) investigation to determine the root cause for why an initial emdr was not previously submitted to the fda in accordance with boston scientific established mdr reporting process and fda adverse event reporting requirements.Information gained through these investigation efforts will be utilized to determine root cause and implement appropriate preventive action(s), as necessary.Analysis of the returned driver revealed that the end of the driver was damaged and both tips of the driver were completely sheared off.The damage to the driver was sufficient to prevent functional testing.The damage to the driver suggests that excessive force was used during the procedure.A review of the instructions for use (ifu) was performed, it states to avoid application of excessive stress on instrumentation and do not force deployment or implant breakage or damage to bony structures may result.
|
| |
|
Additional Manufacturer Narrative
|
|
A retroactive review of events within the quality system identified this record as requiring remediation, including reporting and submission of an initial emdr per applicable criteria.Boston scientific has opened a corrective and preventive action (capa 8647) investigation to determine the root cause for why an initial emdr was not previously submitted to the fda in accordance with boston scientific established mdr reporting process and fda adverse event reporting requirements.Information gained through these investigation efforts will be utilized to determine root cause and implement appropriate preventive action(s), as necessary.Analysis of the returned driver revealed that the end of the driver was damaged and both tips of the driver were completely sheared off.The damage to the driver was sufficient to prevent functional testing.The damage to the driver suggests that excessive force was used during the procedure.A review of the instructions for use (ifu) was performed, it states to avoid application of excessive stress on instrumentation and do not force deployment or implant breakage or damage to bony structures may result.It also states to carefully inspect all instruments prior to and after use.Do not use an instrument that is visibly damaged.If an instrument appears damaged after use confirm that no broken fragments are retained in the patient.
|
| |
|
Event or Problem Description
|
|
It was reported that during the superion indirect decompression system (ids) implant procedure, the driver tip broke.All broken fragments were successfully removed.The broken driver was replaced with a new one and the procedure was completed without further issues.
|
| |
|
Search Alerts/Recalls
|