| Model Number |
102-9800 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/19/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Analysis of the returned driver revealed that the end of the driver was damaged, and both tips of the driver were completely sheared off.The damage to the driver was sufficient to prevent functional testing.The damage to the driver is most likely due to subjecting the device to excessive force due to an unintended error.A review of the instructions for use (ifu) was performed, it states to avoid application of excessive stress on instrumentation and do not force deployment or implant breakage or damage to bony structures may result.
|
| |
|
Event or Problem Description
|
|
It was reported that during the superion indirect decompression system (ids) implant procedure the driver broke, and two end pieces remained in the implant.The physician was ultimately able to retrieve all broken pieces, and a final x-ray confirmed that nothing remained in the patient.The implant procedure was completed with new driver.
|
| |
|
Additional Manufacturer Narrative
|
|
A retroactive review of events within the quality system identified this record as requiring remediation, including reporting and submission of an initial emdr per applicable criteria.Boston scientific has opened a corrective and preventive action (capa 8647) investigation to determine the root cause for why an initial emdr was not previously submitted to the fda in accordance with boston scientific established mdr reporting process and fda adverse event reporting requirements.Information gained through these investigation efforts will be utilized to determine root cause and implement appropriate preventive action(s), as necessary.Fields h7 and h9 were updated.An urgent medical device correction (97003015-fa) was delivered to physicians in may 2023 communicating the potential for superion indirect decompression system (ids) driver instrument tip breaks to occur with use of excessive force during the implant procedure.Driver tip damage may be readily detected via visual, audible, and/or tactile signals.Analysis of the returned driver revealed that the end of the driver was damaged, and both tips of the driver were completely sheared off.The damage to the driver was sufficient to prevent functional testing.The damage to the driver is most likely due to subjecting the device to excessive force due to an unintended error.A review of the instructions for use (ifu) was performed, it states to avoid application of excessive stress on instrumentation and do not force deployment or implant breakage or damage to bony structures may result.
|
| |
|
Event or Problem Description
|
|
It was reported that during the superion indirect decompression system (ids) implant procedure the driver broke, and two end pieces remained in the implant.The physician was ultimately able to retrieve all broken pieces, and a final x-ray confirmed that nothing remained in the patient.The implant procedure was completed with new driver.
|
| |
|
Event or Problem Description
|
|
It was reported that during the superion indirect decompression system (ids) implant procedure the driver broke, and two end pieces remained in the implant.The physician was ultimately able to retrieve all broken pieces, and a final x-ray confirmed that nothing remained in the patient.The implant procedure was completed with new driver.
|
| |
|
Additional Manufacturer Narrative
|
|
A retroactive review of events within the quality system identified this record as requiring remediation, including reporting and submission of an initial emdr per applicable criteria.Boston scientific has opened a corrective and preventive action (capa 8647) investigation to determine the root cause for why an initial emdr was not previously submitted to the fda in accordance with boston scientific established mdr reporting process and fda adverse event reporting requirements.Information gained through these investigation efforts will be utilized to determine root cause and implement appropriate preventive action(s), as necessary.Fields h7 and h9 were updated.An urgent medical device correction (97003015-fa) was delivered to physicians in may 2023 communicating the potential for superion indirect decompression system (ids) driver instrument tip breaks to occur with use of excessive force during the implant procedure.Driver tip damage may be readily detected via visual, audible, and/or tactile signals.Analysis of the returned driver revealed that the end of the driver was damaged, and both tips of the driver were completely sheared off.The damage to the driver was sufficient to prevent functional testing.The damage to the driver is most likely due to subjecting the device to excessive force.Therefore, the most probable cause is due to an unintended use error.A review of the instructions for use (ifu) was performed, it states to avoid application of excessive stress on instrumentation.Additionally, it also states to carefully inspect all instruments prior to and after use.Do not use an instrument that is visibly damaged.If an instrument appears damaged after use, confirm that no broken fragments are retained in the patient.
|
| |
|